Single-Sided Two-Station Glovebox
Model:
Universal(2440-750-900)
Applied range:
Theglove box is widely used in biochemistry, bio-metallurgy, electronics, chemical industry, geology, minerals, medicine, new energy batteries and other electronic materials, magnetic materials, equipment batteries, biological cultivation, sub-packaging and weighing, food vacuum packaging, battery materials and other processes to provide environmental control.
Date issued:
2024-12-19 00:25:37
Equipment Size:
L2440mm* W750mm * H900mm
Weight:
650kg
Details
This glove box is designed for use across a wide range of battery research and production lines, including button cell, prismatic aluminum case, dry electrode preparation, solid-state, single-layer stacked, metal lithium, cylindrical, zinc-ion, pouch cell, and glovebox stacking configurations. Manufactured by Canrd, the universal model (2440-750-900) offers a single-sided, double-station configuration suitable for handling air-sensitive materials during battery assembly. Its robust construction and controlled atmosphere enable precise environmental isolation, critical for processes requiring low moisture and oxygen levels. The 2440 mm length, 750 mm depth, and 900 mm height provide ample working space while maintaining ergonomic access. This glove box supports both fundamental research and scalable production, making it a versatile tool for laboratories working on advanced energy storage technologies.
Transition Compartment Size | Diameter 360mm, Length 600mm, Size Diameter 150mm, Length 300mm |
Equipment Size | L12440mm * W750mm * H900mm |
Site Bearing | / |
Operating Environment | regular Environment |
Power | 1300W |
Power Supply | AC:220V, 50HZ |
Equipment Weight | 650kg |
Fan | 90 m/h |
Oxygen Analyzer | Measurement Range: 0 ~ 1000ppm |
Water Analyzer | Measurement Range: 0 ~ 500ppm |
Water Index | <1PPM |
Vacuum Pump | 8 m/h |
Vacuum Pump Limit Vacuum | ≤ 2x 10 -1pa |
Oxygen Index | <1PPM |
& nbsp; & nbsp; & nbsp; The glove box is widely used in biochemistry
- closed circulation: the inert gas in the glove box is closed circulation through the circulation fan and purifier to continuously remove water and oxygen;
- Automatic control regeneration: the water and oxygen removal material can be regenerated, and the regeneration process is controlled by the program;
- Automatic cleaning: The atmosphere replacement in the glove box is completed by an automatically controlled cleaning valve;
- box pressure control: the pressure in the glove box is automatically controlled by PLC, which can be set freely, and the system is automatically protected beyond +/- 15mbar;
- Vacuum pump automatic control: The vacuum pump is required to be automatically turned on when the system needs it.
R&D 지원이 필요하신가요?
주요 제품 카테고리를 확인하고 배터리 소재를 Canrd R&D 서비스와 연결하세요.
실험 소재
카테고리배터리 연구용 양극, 음극, 전해액, 분리막 및 기타 소재.
양극, 음극 및 전해액 소재 보기 →실험 장비
카테고리파일럿 라인, 지그 및 테스트 워크플로용 배터리 실험 장비.
배터리 실험 장비 보기 →전극 및 셀
카테고리R&D 검증을 위한 맞춤 전극, 드라이 셀 및 배터리 셀 부품.
맞춤 배터리 전극 및 셀 보기 →나트륨 전해액
카테고리나트륨 배터리 연구 및 검증을 위한 나트륨 이온 전해액 옵션.
나트륨 배터리 전해액 보기 →R&D 서비스
서비스소재 평가부터 셀 제작 및 고장 분석까지 전주기 배터리 R&D 서비스.
Canrd R&D 서비스 살펴보기 →소재 평가
서비스배터리 소재의 특성 분석, 전기화학 테스트 및 호환성 평가.
배터리 소재 평가하기 →전극 제작
서비스목표 화학계와 공정 요구에 맞춘 맞춤 전극 시트 및 롤.
전극 제작 요청하기 →셀 제작
서비스R&D 검증을 위한 드라이 셀, 조립 셀 및 맞춤 셀 형식.
맞춤 연구 셀 제작하기 →문의
페이지Canrd가 프로젝트에 맞는 제품, 소재 및 서비스를 연결해 드립니다.
제품 및 R&D 지원 받기 →