Details
Sulfide Electrolyte Powder Ionic Conductivity Test:
1. The ionic conductivity of sulfide solid electrolyte powder increases with increasing tableting pressure. Above 400 MPa, the ionic conductivity tends to stabilize. Excessive pressure may even cause internal cracking and delamination of the tablet, resulting in inaccurate test values. Therefore, tableting is generally performed at a pressure of 200-400 MPa.
2. A stainless steel post is used as a blocking electrode in the cold-pressed green sheet. Applying a pressure of at least 30 MPa on both sides ensures good contact between the electrode and the electrolyte.
3. Testing Procedure:
① Tableting: Pour 200 mg of electrolyte powder into a mold with an inner diameter of 10 mm and apply a pressure of 350 MPa to the tablet.
② Measurement: Use a vernier caliper depth gauge to measure the thickness of the electrolyte sheet in the mold.
③ Assembly: After assembly, apply 350 MPa of pressure, maintain pressure using a steel pressure gauge, and then test.
④ EIS Test: Connect to the electrochemical workstation, set the test parameters, and start the test. The AC impedance test frequency range is 1 MHz to 1 Hz, with a voltage amplitude of 10 mV.
⑤ Calculation of Ionic Conductivity: Because sulfide materials are soft and have good contact between particles, the AC impedance curve appears as a straight line inclined approximately 45° upward, with no obvious grain boundary capacitance signal. The intersection of the curve and the real axis Z' is used as the total impedance for calculating ionic conductivity.
Important Notes and Usage Recommendations:
1. Operating Environment: All operations must be performed in a glove box filled with high-purity argon (O₂ & H₂O < 0.1 ppm). Any exposure to air will cause material decomposition, performance failure, and the generation of toxic gases (H₂S).
2. Personal Protection Equipment (PPE): Even when working in a glove box, it is recommended to wear chemical-resistant gloves (such as nitrile rubber gloves). If removal is required, a protective mask, goggles, and an anti-static lab coat must be worn.
Hazard Statements:
Contact with water may release hydrogen sulfide gas, which is colorless and highly toxic. It has a rotten egg odor at low concentrations and is odorless at high concentrations. It may spontaneously ignite.
2. Fatal if swallowed, in contact with skin, or inhaled.
3. Can cause severe skin burns and eye damage.
4. Extremely toxic to aquatic life.
Precautionary Instructions:
1. Avoid contact with water.
2. Operate in an inert gas glove box to protect from moisture.
3. Wear protective gloves/protective clothing/eye protection/face protection.
4. Wash thoroughly after handling.
5. Do not eat, drink, or smoke when using this product.
6. Do not breathe dust/fume/gas/mist/vapor/spray.
7. Avoid release to the environment.
Safe Storage:
1. This product is packaged in a fluorinated bottle in a glove box, sealed magnetically, and then sealed in a double aluminum-plastic bag. It can be stored under normal conditions if unopened.
2. If opened for use, it must be opened and stored in a glove box under an inert atmosphere (e.g., argon) with a water-oxygen concentration of less than 0.1 ppm.
Emergency Measures:
1. Ingestion: Immediately call a poison control center or doctor. Rinse mouth. Do not induce vomiting.
2. Skin (or hair) contact: Immediately remove all contaminated clothing. Rinse skin with plenty of water/shower. Get medical advice/attention immediately.
3. Inhalation: Remove to fresh air and keep comfortable for breathing. Get medical advice/attention immediately.
4. If in eyes: Rinse cautiously with water for several minutes. Remove contact lenses, if present and easy to do. Continue rinsing. Get medical advice/attention immediately.
5. Use a Class D fire extinguisher (for metal fires) or dry sand. Do not use water, foam, or carbon dioxide to extinguish fires.
Disposal:
1. Dispose of waste as hazardous chemicals. Contact a qualified recycling agency for recycling. Do not discard it at will.
2. Lightly contaminated items can be cleaned in a fume hood using an alkaline solution.

XRD

SEM

Delivery time of 1-2 weeks, can consult customer service
The product must be unpacked inside a glove box. In air, it may release hydrogen sulfide (a toxic gas).
It can be shipped overseas, but please contact us in advance to confirm the destination country.

Test conditions:
Take about 0.2g of powder and press it into 10-15MPa tablets (equivalent to a compression of about 250MPa), measure the thickness and diameter;
Place the carbon coated aluminum foil on both sides and form a blocking electrode in the coin cell battery/testing mold for impedance testing;
The powder dosage is adjusted according to the diameter and thickness.